Optical Scintillation Imaging for Dynamic Source Tracking Dosimetry During Prostate HDR Brachytherapy

arXiv Physics
CC BY
이 매체는 공공·자유 라이선스로 본문을 직접 표시합니다.

Abstract

High-dose-rate (HDR) brachytherapy is highly conformal, but its accuracy depends on catheter and source placement, and clinical workflows lack real-time in vivo dosimetry (IVD) that verifies delivery without disrupting the procedure.

We characterize a scintillator-based optical dosimeter integrated into a transrectal probe that tracks source position, speed, dose rate, and time-resolved cumulative dose.

A 67 mm x 5 mm line scintillator is imaged by a CMOS camera through a 30-degree endoscopic lens; geometric and flat-field corrections remove barrel distortion and vignetting.

Absolute dose was calibrated with a TOPAS Monte Carlo model of the probe's air-cavity geometry.

The system was validated in a water phantom with an Ir-192 source and needles at radial offsets of 7, 10, and 15 mm.

Response varied by $\pm$2\% along the 67 mm scintillator and was linear over 5-21 cGy/s ($R^2 \geq 0.995$).

The Monte Carlo model showed a 3.75\% dose-rate reduction relative to TG-43 water-equivalent conditions.

In phantom, the system resolved a 3 mm needle displacement, confirmed by CT.

Measured dose-rate profiles agreed with Monte Carlo in peak magnitude and spatial distribution, and integral dose agreed within 1\%.

By mapping dose directly to the measurement plane of a standard ultrasound transducer, this system provides high-resolution, time-resolved verification of HDR delivery and a clinical tool for detecting delivery errors with minimal changes to existing treatment protocols.