Monte Carlo-based secondary dose check platform for the world's first ultra-compact upright proton therapy system
Abstract
Objective.
This work aimed to develop and validate an independent Monte Carlo (MC) secondary dose check of the RayStation treatment planning system (TPS) for the Mevion S250-FIT, the world's first ultra-compact upright proton therapy system.
Approach.
The treatment nozzle, including the energy modulation system of boron carbide and polycarbonate plates and the adaptive aperture, was modeled in the TOPAS (TOol for PArticle Simulation) MC code, and the beam model was commissioned against integrated depth dose curves, in-air spot profiles, and dose per monitor unit measured at 12 energy settings.
Based on this model, an in-house converter, RayStation2TOPAS, recalculates the TPS plans by reconstructing each pencil beam spot, the aperture for each energy layer, and the seated patient's rotation.
In total, 15 water phantom plans and 25 clinical plans were recalculated and compared with the TPS by the point dose, with a tolerance of 5%, and by three-dimensional gamma analysis at 3%/2mm, with a pass rate of at least 90%.
Main results.
The simulated proton range agreed with the measurement within 0.25mm, the spot full width at half maximum within 6.8%, and the dose per monitor unit within 1.3%.
The point dose differed from the TPS by -0.76% to +1.98% in the water phantom plans and by -3.03% to +3.99% in the clinical plans, with gamma pass rates of 96.39% to 99.93% and 90.08% to 99.98%, so every plan passed both tests.
Significance.
To our knowledge, this is the first independent MC secondary dose check for the S250-FIT, and it complements measurement-based patient-specific quality assurance of its plans.