Pressure behavior in a roller pump-driven low-flow VACC mock circuit with inactive de-airing pump: A descriptive bench characterization under reduced patient-side effective filling pressure conditions
Figures
Abstract
This exploratory bench study characterized pressure behavior in a roller pump-driven low-flow valve-type semi-closed extracorporeal circulation (VACC) mock circuit under experimentally reduced patient-side effective filling pressure conditions with the de-airing pump inactive. Pressures were measured at four sites: circuit-reservoir pressure (Pc), patient-side mock-reservoir outlet pressure (Pv), roller pump inlet pressure (Pi), and roller pump outlet pressure (Po). Pump flow-rate settings were 50–300 mL/min. Circuit-reservoir negative-pressure setpoints were 0, −10, −20, −30, and −40 mmHg, and patient-side mock-reservoir negative-pressure setpoints were 0, −5, and −10 mmHg. A 32 wt% aqueous glycerol solution was used. Pc generally reflected the circuit-reservoir negative-pressure setpoint, with a mean difference of −0.18 mmHg (95% CI, −0.37 to 0.00). The 95% CIs summarize variability among technical replicates within this bench system and should not be interpreted as population-inferential estimates. Pv decreased stepwise according to the patient-side setpoint. Within the common evaluation range, mean differences versus 0 mmHg were −5.7 mmHg (95% CI, −6.0 to −5.3) for the −5 mmHg condition and −9.2 mmHg (95% CI, −9.5 to −8.9) for the −10 mmHg condition. Pi shifted in the negative direction as the pump flow-rate setting increased, whereas Po increased monotonically under the fixed occluder condition. Delivered flow was independently verified volumetrically before the main experiments at 100, 200, and 300 mL/min but was not continuously measured during the pressure experiments. Therefore, Po was interpreted primarily in relation to the pump flow-rate setting rather than as a directly measured hydraulic pressure–flow relationship. These data provide baseline pressure-characterization information for future mock-circuit studies and do not establish clinical safety, efficacy, or a stable operating range.
Citation: Wauchi K, Okumura T, Matsuda K, Sato H, Miyata M, Tani K (2026) Pressure behavior in a roller pump-driven low-flow VACC mock circuit with inactive de-airing pump: A descriptive bench characterization under reduced patient-side effective filling pressure conditions. PLoS One 21(11): e0363658. https://doi.org/10.1371/journal.pone.0363658
Editor: Fatemeh Afsharnia, Ferdowsi University of Mashhad, IRAN, ISLAMIC REPUBLIC OF
Received: July 24, 2026; Accepted: November 10, 2026; Published: November 25, 2026
Copyright: © 2026 Wauchi et al. This is an open access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
Data Availability: The replicate-level experimental dataset and analysis code used to reproduce the statistical analyses, summary tables, and Figs 2–4 and S1 and S2 Figs are provided as Supporting Information. The condition-specific summary statistics are provided as S1 Table, and the full regression results for Po are provided as S2 Table. Additional pre-experimental verification data, including volumetric pump-flow calibration raw data and raw fixed flow rate-injection pressure (FIP) data used to document the roller-pump occlusion setting, are publicly available from Zenodo (Version v2, DOI: 10.5281/zenodo.22604656).
Funding: The author(s) received no specific funding for this work.
Competing interests: The authors have declared that no competing interests exist.
Abbreviations: CI, Confidence interval; CPB, Cardiopulmonary bypass; D-pump, De-airing pump; ECC, Extracorporeal circulation; ECMO, Extracorporeal membrane oxygenation; FIP, Fixed flow rate-injection pressure; JIS, Japanese Industrial Standards; ND, Not determined; Pc, Circuit-reservoir pressure; Pi, Roller pump inlet pressure; Po, Roller pump outlet pressure; Prc,set, Circuit-reservoir negative-pressure setpoint; Prp,set, Patient-side mock-reservoir negative-pressure setpoint; Pv, Patient-side mock-reservoir outlet pressure; SD, Standard deviation; VACC, Valve-type semi-closed extracorporeal circulation
Introduction
Cardiac surgery in small and toy-breed dogs presents technical challenges in cardiopulmonary bypass (CPB) management because low-flow conditions are associated with small body size, limited blood volume, and the relatively large priming volume of the circuit [1,2]. In low-body-weight dogs, slight variations in circuit volume, cannula diameter, venous drainage conditions, or pump conditions may affect circulatory support. The selection of arterial cannulas may also be limited in toy-breed dogs weighing <3 kg [3]. Therefore, optimization of patient-specific factors, circuit configuration, priming volume, venous drainage, and arterial flow conditions remains an important technical issue in veterinary CPB [4,5]. In this context, characterization of circuit pressure behavior in the low-flow range is a necessary preliminary step before evaluating broader operational performance or clinical applicability.
In closed circuits without a reservoir, as typified by extracorporeal membrane oxygenation (ECMO) support, extracorporeal blood flow is regulated by the main pump [6]. In contrast, conventional open CPB circuits require separate adjustment of arterial flow and venous drainage while the reservoir level is managed. In low-flow circuits with limited circulating volume, even a slight imbalance between arterial flow and venous drainage may alter the reservoir level. Such changes may require careful coordinated control [4,5]. When the concept of a closed circuit is applied to CPB management, the circuit must accommodate both additional inflow through surgical suction or vent return and insufficient supply from the patient side to the venous drainage line. A simple closed circuit has limited ability to passively buffer such volume fluctuations via a reservoir. Valve-type semi-closed extracorporeal circulation (VACC) is based on a conventional open-type ECC circuit and incorporates a one-way valve at the reservoir outlet, thereby combining closed-circuit behavior with the reservoir-mediated volume buffering of an open circuit [7]. One reported design concept of VACC is to simplify extracorporeal circulation management by combining a one-way valve with a circuit reservoir [8]. The present study did not evaluate the safety or clinical performance of this concept. In the VACC concept, the circuit reservoir functions as a volume-control reservoir and provides tolerance to volume fluctuations when additional inflow or insufficient venous drainage-side supply occurs. Specifically, when pressure on the venous drainage side decreases excessively relative to reservoir pressure, the one-way valve opens and allows fluid to be supplied from the reservoir to the circuit. Reservoir pressure is therefore an important factor in understanding the fundamental behavior of VACC [7,9].
VACC was developed as a centrifugal pump-driven CPB circuit for adult human patients, and its clinical implementation has been reported [7,8]. However, CPB for low-body-weight patients differs from adult CPB in terms of available devices, circuit internal diameter, pump specifications, and flow range. Therefore, the behavior of VACC reported in adult patients and centrifugal pump-driven systems cannot be directly extrapolated to a roller pump-driven low-flow circuit designed for small dogs weighing ≤3 kg. Before the operational performance of a roller pump-driven low-flow VACC configuration can be evaluated, its fundamental pressure behavior under controlled small-bore mock-circuit conditions should be characterized.
In a low-flow VACC circuit using a roller pump, roller pump inlet pressure, arterial-side boundary conditions, patient-side effective filling pressure, and circuit-reservoir negative pressure may interact. Particularly under conditions simulating small dogs weighing ≤3 kg, pump flow-rate settings are as low as 50–300 mL/min. Pressure changes and hydrostatic pressure effects that may be relatively minor in adult circuits could therefore substantially affect the interpretation of circuit behavior. Impaired venous drainage and venous collapse have been shown in a bench model to contribute to flow limitation during CPB [10]. In the present study, the circuit-reservoir negative-pressure setpoint (Prc,set) and the patient-side mock-reservoir negative-pressure setpoint (Prp,set) were defined as operational factors for distinct purposes. Prc,set was used as a circuit-side boundary condition intended to regulate circuit-reservoir pressure (Pc). The one-way valve was not directly instrumented; therefore, its state was not quantitatively determined. In contrast, Prp,set was used to reduce patient-side effective filling pressure and simulate impaired venous drainage. Experimental systems involving negative pressure require attention to negative-pressure settings, air entrainment, microbubbles, hemolysis, and reservoir-pressure management [11–13]. However, negative-pressure application in the present study was used experimentally to establish circuit-side and patient-side pressure boundary conditions. It was not used to evaluate negative pressure as a venous drainage assistance method. Nevertheless, fundamental data systematically characterizing Pc, patient-side mock-reservoir outlet pressure (Pv), roller pump inlet pressure (Pi), and roller pump outlet pressure (Po) under standardized reductions in patient-side effective filling pressure remain limited. Therefore, as an initial step, pressure behavior should be descriptively characterized under conditions without de-airing pump (D-pump)-mediated de-airing or fluid return. In the VACC circuit, the D-pump is used to remove air captured in the bubble trap and to return fluid to the circuit reservoir. This characterization should include whether Prc,set is reflected in Pc and how reduced patient-side effective filling pressure conditions are reflected in Pv, Pi, and Po. Because the mock circuit did not reproduce biological circulation or clinical reservoir management, the present analysis was limited to pressure behavior under standardized mock-circuit conditions. It was not intended to determine clinical safety, clinical efficacy, the stable operating range of VACC, actual circuit flow, venous drainage adequacy, reservoir-level control, one-way valve dynamics, air-handling performance, hemolysis, or D-pump-assisted operation.
In this study, the term “reduced patient-side effective filling pressure” does not denote physiological venous pressure. It denotes an experimentally imposed reduction in the pressure available at the patient-side mock-reservoir outlet under the defined hydrostatic and circuit-height conditions. This condition was empirically represented by measured Pv. The aim of this study was to descriptively characterize the pressure behavior of Pc, Pv, Pi, and Po in a low-flow, roller pump-driven VACC mock circuit under standardized bench conditions. Specifically, we examined the effects of reduced patient-side effective filling pressure conditions and Prc,set on each pressure measurement site. Prp,set was used as an operational factor, and measured Pv was treated as its empirical pressure indicator. Pc was used to confirm the reflection of Prc,set conditions. Pi was used as a supportive indicator of roller pump inlet-side load. Po was used to characterize its dependence on the pump flow-rate setting under standardized arterial-side boundary conditions. Previous VACC reports primarily addressed the development and clinical implementation of adult centrifugal pump-driven systems [7,8]. Previous bench work on impaired venous drainage focused on phenomena such as venous collapse rather than pressure characterization of a low-flow VACC configuration [10]. In contrast, the present study examined a small-bore, roller pump-driven low-flow VACC mock circuit. The circuit-reservoir negative-pressure setpoint, patient-side mock-reservoir negative-pressure setpoint, and pump flow-rate setting were separately manipulated while Pc, Pv, Pi, and Po were measured. The novelty of this study therefore lies in the systematic separation of these operational setpoints from the corresponding measured pressure responses. In particular, the study distinguishes Prp,set as an experimental operating condition from Pv as a measured indicator of reduced patient-side effective filling pressure conditions.
Materials and methods
Study design
This exploratory bench study evaluated pressure behavior in a low-flow, roller pump-driven VACC mock circuit configured to represent a small-dog-scale low-flow experimental setting. The operational factors were Prp,set, Prc,set, and the pump flow-rate setting. Correspondence between the pump flow-rate setting and delivered flow was independently verified volumetrically before the main experiments at 100, 200, and 300 mL/min. However, delivered flow was not continuously measured during the pressure experiments. Therefore, the pump flow-rate setting was retained as the operational flow variable in the primary analysis. In this bench model, the term “reduced patient-side effective filling pressure” did not represent physiological venous pressure, central venous pressure, or in vivo venous drainage pressure. This condition was operationally created by applying Prp,set to the patient-side mock reservoir and was empirically assessed using measured Pv. The outcome measures were the pressure behaviors of Pc, Pv, Pi, and Po. The measurement sites and roles of these pressure variables are summarized in Table 1. The primary descriptive endpoints were (1) agreement between Prc,set and measured Pc, (2) stratification of measured Pv according to Prp,set, and (3) the evaluable pump flow-rate setting range before the prespecified termination criteria were reached. The secondary descriptive endpoints were the responses of Pi and Po to increasing pump flow-rate settings under the fixed arterial-side boundary condition.
Repeated measurements were technical replicates obtained using the same bench circuit. Accordingly, this study was not intended to support confirmatory inference to a clinical population. Pressure behavior across conditions was evaluated descriptively and using estimation-based methods. This study was designed as an initial pressure-characterization step rather than as an evaluation of complete VACC performance, clinical safety, or stable operating range.
Experimental circuit
The mock experimental setup was constructed on the basis of a VACC circuit. A patient-side mock reservoir was incorporated to create reduced patient-side effective filling pressure conditions. The setup comprised a circuit reservoir, an oxygenator, a patient-side mock reservoir, a bubble trap, a de-airing pump (D-pump), a roller pump, a one-way valve, an arterial-line occluder, and four pressure measurement sites. The overall circuit configuration is shown in Fig 1. The pressure measurement sites were defined as the circuit-reservoir pressure (Pc), patient-side mock-reservoir outlet pressure (Pv), roller pump inlet pressure (Pi), and roller pump outlet pressure (Po).
The mock circuit was configured as a closed-circuit-like main flow path with a side-branch circuit reservoir connected through a one-way valve. In the VACC concept, the circuit reservoir functions as a volume-control reservoir, and the one-way valve is intended to provide a reservoir-to-circuit fluid path when drainage-side pressure decreases relative to reservoir pressure. Pc, Pv, Pi, and Po denote the circuit-reservoir pressure, patient-side mock-reservoir outlet pressure, roller pump inlet pressure, and roller pump outlet pressure, respectively. Prc,set and Prp,set denote the circuit-reservoir negative-pressure setpoint and patient-side mock-reservoir negative-pressure setpoint, respectively. The pressure transducers for Pc, Pv, Pi, and Po were zero-balanced at the reference height of the main roller pump, and the outlet of the patient-side mock reservoir was fixed 10 cm above this reference height. The initial working-fluid volumes in the circuit reservoir and patient-side mock reservoir were 100 and 200 mL, respectively. The roller pumps used for negative-pressure application were experimental devices used to establish Prc,set and Prp,set and were not components of the clinical VACC circuit. The D-pump, shown as the de-airing and fluid-return path associated with the bubble trap, remained inactive during the primary analysis.
A Stockert S5 CPB system (LivaNova Japan K.K., Tokyo, Japan) was used. The roller pump, D-pump, and arterial-line occluder were components integrated into the Stockert S5 system. A CAPIOX® FX25 membrane oxygenator with an integrated venous reservoir (Terumo Corporation, Tokyo, Japan) was used as the circuit reservoir and oxygenator. A CX-RR10 reservoir (Terumo Corporation, Tokyo, Japan) was used as the patient-side mock reservoir. A MERA Bubble Trap HBT-C1 (Senko Medical Instrument Mfg. Co., Ltd., Tokyo, Japan) was used as the bubble trap. A RetroGuard® Arterial Check Valve (Quest Medical, Inc., Allen, TX, USA), a duckbill-type one-way valve, was used as the one-way valve.
The D-pump was an auxiliary de-airing pump used to remove air captured in the bubble trap and return fluid to the circuit reservoir. The bubble trap and D-pump were incorporated to address air that may enter from the venous drainage side in closed or semi-closed circuits. To evaluate baseline pressure behavior without de-airing or associated fluid return, the D-pump remained inactive during the primary analysis.
Exceline Tube H tubing (1/4 × 3/32 inch; Senko Medical Instrument Mfg. Co., Ltd., Tokyo, Japan) was used for the main circuit. The tubing had an internal diameter of 1/4 inch (6.35 mm) and a wall thickness of 3/32 inch (2.38 mm). The outer diameter calculated from the internal diameter and wall thickness was 7/16 inch (approximately 11.1 mm). Tubing of the same specification was used in the main-circuit roller pump head. The main-circuit roller pump was equipped with an 85-mm-diameter pump head in the Stockert S5 system. Exceline Tube H tubing (3/8 × 3/32 inch; Senko Medical Instrument Mfg. Co., Ltd., Tokyo, Japan) was used in the pump-head segments of the roller pumps used for negative-pressure application. This tubing had an internal diameter of 3/8 inch (9.53 mm) and a wall thickness of 3/32 inch (2.38 mm). The outer diameter calculated from the internal diameter and wall thickness was 9/16 inch (approximately 14.3 mm). Each roller pump used for negative-pressure application was equipped with a 150-mm-diameter pump head in the Stockert S5 system. Except for the pump-head segments, the negative-pressure application lines consisted of the same 1/4 × 3/32 inch tubing used in the main circuit. Four pressure-monitoring tubing sets (PXMKP11152, Edwards Lifesciences K.K., Tokyo, Japan), each incorporating a pressure transducer, were connected to two BSM-2303 physiological monitors (Nihon Kohden Corporation, Tokyo, Japan). These monitors were used to measure Pc, Pv, Pi, and Po. According to the manufacturer specifications for the BSM-2303, the invasive-pressure measurement range was −50–300 mmHg. Measurement accuracy was ± 1 mmHg ± 1 digit from −50 to <100 mmHg and ±1% ± 1 digit from 100 to 300 mmHg. The auto-zero accuracy was ± 1 mmHg [14].
At the beginning of each measurement condition, the working-fluid volume in the VACC circuit reservoir was adjusted to 100 mL. The working-fluid volume in the patient-side mock reservoir was adjusted to 200 mL. The outlet of the patient-side mock reservoir was fixed at 10 cm above the reference height of the main roller pump. The outlet, rather than the fluid surface, was used as the fixed geometric reference because the fluid level in the patient-side mock reservoir could vary during testing. The pressure transducers for Pc, Pv, Pi, and Po were zero-balanced at the reference height of the main roller pump. Accordingly, the pressure values reported in this study were recorded as gauge pressures relative to the reference height of the main roller pump. They were not referenced to the local hydrostatic head at each measurement site. Based on the working-fluid density of approximately 1.076 g/mL at 25°C, a 10-cm vertical column of the working fluid corresponds to a hydrostatic pressure difference of approximately 7.9 mmHg. However, the instantaneous fluid-surface height in the patient-side mock reservoir varied during testing and was not recorded separately. Therefore, the total hydrostatic contribution to measured Pv could not be quantitatively decomposed from the imposed Prp,set and other circuit effects.
A 32 wt% aqueous glycerol solution was used as the working fluid. Aqueous glycerol solutions have been used as working fluids to approximate blood viscosity in mock circulation circuits [15]. The viscosity of the working fluid was measured at 25°C using an SV-10 tuning-fork vibration viscometer (A&D Company, Tokyo, Japan). The measured viscosity was 2.80 ± 0.05 cP (n = 6), which was within the reported range of whole-blood viscosity under hemodilution conditions [16]. The density corresponding to a 32 wt% aqueous glycerol solution at 25°C was approximately 1.076 g/mL, based on published density data for aqueous glycerol solutions [17]. However, the solution is approximately Newtonian and does not reproduce cellular or non-Newtonian blood properties, including shear dependence, red blood cell deformability, and aggregation [18]. Accordingly, this simplified model was used to evaluate pressure behavior while excluding the effects of cellular components and non-Newtonian behavior.
An HHC-51 temperature control unit (Senko Medical Instrument Mfg. Co., Ltd., Tokyo, Japan) was connected to the heat exchanger of the oxygenator. This was done to match the working-fluid temperature during the experiments to the temperature at which viscosity was measured. The working-fluid temperature within the circuit was maintained at 25.0 ± 0.5°C. The fluid temperature was verified using the monitor display of the Stockert S5 system. The temperature of 25°C was selected as a temperature-control condition for this bench study and was not intended to reproduce clinical body temperature or any specific hypothermia-management condition.
The lengths of the venous drainage line and arterial line were standardized to 100 cm. Pressure at the patient-side mock-reservoir outlet was affected by the hydrostatic pressure associated with the fluid level in the reservoir. Therefore, measured Pv, rather than Prp,set alone, was treated as an empirical indicator of reduced patient-side effective filling pressure.
Before testing, the degree of roller pump occlusion was adjusted with the 1/4-inch pump tubing mounted in the 85-mm-diameter pump head of the Stockert S5 system. The initial occlusion adjustment was based on the concept of appropriate occlusion described in JIS T 1603 [19]. The degree of occlusion was subsequently assessed using the previously reported fixed flow rate-injection pressure (FIP) method [20]. During FIP assessment, the same 32 wt% aqueous glycerol solution was injected at a constant rate of 0.50 mL/min (30 mL/h) using a syringe pump, and the FIP pressure was recorded for 300 s. The FIP pressure at 300 s was 53 mmHg. This value was used to quantitatively document the roller-pump occlusion setting employed in the present experiment. It was not interpreted as a universally validated optimal threshold for the specific combination of 1/4-inch tubing and an 85-mm-diameter pump head. The degree of roller-pump occlusion was not altered during testing.
Experimental conditions
The pump flow-rate setting was adjusted to six levels: 50, 100, 150, 200, 250, and 300 mL/min. This range was selected to include low-flow conditions corresponding to a maximum of approximately 100 mL/kg/min in the 3-kg model [1,2]. The phrase “designed for small dogs weighing ≤3 kg” refers to the intended low-flow range and circuit scale of the mock setup. It does not indicate that the system was validated for clinical use in dogs of this body weight. Before the main pressure experiments, after adjustment of roller-pump occlusion, correspondence between the pump flow-rate setting and delivered flow was independently verified using a volumetric collection method. Verification was performed at pump flow-rate settings of 100, 200, and 300 mL/min under the same circuit conditions using the same 32 wt% aqueous glycerol solution at approximately 25°C. Fluid delivered over 60 s was collected using three 100-mL graduated cylinders, and six repeated measurements were obtained at each pump flow-rate setting. Mean measured flows were 100.0 ± 0.6, 200.7 ± 0.8, and 300.8 ± 0.8 mL/min at pump flow-rate settings of 100, 200, and 300 mL/min, respectively. Delivered flow was not continuously measured during the subsequent pressure experiments.
Prc,set and Prp,set were established as experimental conditions for the circuit reservoir and the patient-side mock reservoir, respectively. Prc,set was defined as the negative-pressure setpoint applied to the circuit reservoir. The Prc,set conditions were 0, −10, −20, −30, and −40 mmHg. The lower measurement limit of the monitor used in this study was −50 mmHg. Therefore, the lower limit of the experimental negative-pressure range was set at −40 mmHg to ensure an adequate margin within the measurable range.
Prp,set was defined as the negative-pressure setpoint applied to the patient-side mock reservoir and was used as an operational factor to establish reduced patient-side effective filling pressure conditions. The Prp,set conditions were 0, −5, and −10 mmHg. However, Pv, measured at the outlet of the patient-side mock reservoir, was affected by Prp,set as well as by the outlet height of the patient-side mock reservoir and the corresponding hydrostatic head. Therefore, measured Pv was used as the empirical indicator of reduced patient-side effective filling pressure conditions in the analysis. All pressure values were recorded as gauge pressures relative to the same reference height using pressure transducers zero-balanced at the reference height of the main roller pump.
Prc,set and Prp,set were established using two independent experimental negative-pressure application systems, one for the circuit reservoir and one for the patient-side mock reservoir. Each system used a roller pump equipped with a 150-mm-diameter pump head of the Stockert S5. Exceline Tube H tubing (3/8 × 3/32 inch) was used in the roller pump head segments of the two negative-pressure application systems, whereas 1/4 × 3/32 inch tubing was used for the remaining segments of the negative-pressure application lines. The negative-pressure application lines were connected to the air phase above the working fluid in the circuit reservoir and the patient-side mock reservoir. This configuration applied negative pressure to the air phase without directly aspirating the working fluid from either reservoir.
Two identical negative-pressure control systems were used, one for the circuit reservoir and one for the patient-side mock reservoir. In both systems, the source vacuum indicated by the high-range gauge was maintained between −400 and −600 mmHg. For the circuit-reservoir system, the regulated downstream negative pressure was adjusted using the low-range gauge to the predefined Prc,set values of 0, −10, −20, −30, and −40 mmHg. For the patient-side mock-reservoir system, it was adjusted to the predefined Prp,set values of 0, −5, and −10 mmHg. Prc,set and Prp,set were adjusted according to the negative-pressure values indicated by the respective controllers, independently of the Pc and Pv measurements obtained using the BSM-2303 physiological monitors. The negative-pressure application pumps were operated continuously during measurements, and the applied negative pressures were manually fine-adjusted as necessary to maintain the respective target setpoints. Measurements were initiated after confirming that the controller-indicated setpoints and the corresponding measured pressures had stabilized.
An occluder integrated into the Stockert S5 was placed in the arterial line and maintained at a fixed displayed setting to provide a standardized arterial-side boundary condition. The occluder setting was determined in a preliminary test with the arterial-line tubing installed. During the preliminary test, the occluder setting was adjusted stepwise. It was confirmed that Po did not show sustained negative pressure over the range of 50–300 mL/min. Po also did not exceed the upper measurement limit of the monitor, 300 mmHg, at 300 mL/min. Based on these observations, the displayed occluder setting was fixed at 25%. This displayed value did not represent the actual percentage reduction in the luminal cross-sectional area. Instead, it was treated as a fixed device setting used to standardize the arterial-side boundary condition in the present experimental setup. The occluder setting was not changed during the experiments.
Outcome measures and data processing
Within each combination of Prp,set and Prc,set, measurements were performed by sequentially increasing the pump flow-rate setting from 50 to 300 mL/min. When an experimental condition was changed, Prc,set, Prp,set, and the pump flow-rate setting were readjusted to the prespecified values. The working-fluid volumes in the reservoirs were also readjusted to the prespecified values. After pump adjustment, a stabilization period of 60 s was allowed. The maximum, minimum, and mean values of Pc, Pv, Pi, and Po displayed on the two BSM-2303 physiological monitors were then recorded. The order of Prp,set and Prc,set condition combinations was randomized before testing. Within each condition combination, pump flow-rate settings were evaluated in ascending order to apply the prespecified measurement termination criteria consistently. Reservoir-level changes, potential air entrainment into the venous drainage line, and obvious backflow were assessed visually by two experimenters (K.W. and Chisato Ikubo). These observations were used as experimental condition checks and to apply the termination criteria, rather than as blinded primary outcome measures. No formal inter-observer adjudication rule or agreement metric was prespecified for these visual assessments. Therefore, termination status was interpreted as an experimental condition outcome within this bench setup, not as an objective clinical failure endpoint.
The monitor-displayed mean values were used in the primary analysis as representative values of the steady-state pressure conditions. The maximum and minimum values were recorded to confirm the range of displayed pressure fluctuations but were not included in the primary analysis. Continuous pressure waveforms were not stored using an external data logger. Therefore, the monitor-displayed mean values could not be recalculated from externally recorded waveforms. Instantaneous pressure fluctuations, pulsatile components, and frequency components arising from roller pump operation were outside the scope of the analysis.
Pressure values were recorded from the monitor-displayed mean values after the 60-s stabilization period. No additional temporal averaging or offline post-processing of pressure values was performed because continuous pressure waveforms were not externally recorded. The manufacturer documentation reviewed did not specify a separate display-resolution value. Therefore, the recording granularity of the pressure values used for analysis is reported instead. Pressure values were recorded in whole-mmHg increments. Therefore, condition-specific variability, including SD values of 0.00 mmHg, should be interpreted in light of the recording resolution and measurement accuracy of the pressure-monitoring system. An SD of 0.00 mmHg indicates no variation among technical replicates at the recorded resolution and does not indicate the absence of instantaneous pressure fluctuations.
Six technical replicate runs were performed for each Prp,set–Prc,set condition. In each replicate run, the pump flow-rate setting was sequentially increased from 50 mL/min to the maximum evaluable setting for that condition. After a 60-s stabilization period at each flow-rate setting, the monitor-displayed mean values of Pc, Pv, Pi, and Po were recorded. After completion of each replicate run, the pump flow-rate setting was returned to 50 mL/min. The working-fluid volumes in the VACC circuit reservoir and patient-side mock reservoir were readjusted to 100 and 200 mL, respectively. Prc,set and Prp,set were also re-established at their prespecified values before the next replicate run was initiated. These replicate runs were performed using the same circuit and working fluid to assess technical repeatability and did not represent independent biological samples [21]. Accordingly, the means, standard deviations, and 95% confidence intervals for each condition were treated as descriptive indicators of variability among technical replicate measurements within this bench system.
Because the de-airing procedure and associated fluid return resulting from D-pump operation were outside the scope of this study, pressure behavior of Pc, Pv, Pi, and Po was evaluated with the D-pump inactive. Changes in the reservoir level were assessed by visual inspection.
Before testing and during measurements, fluid movement in the patient-side mock reservoir, venous drainage line, and area surrounding the oxygenator was visually inspected. Video recordings were obtained when necessary to confirm the absence of obvious backflow or air entrainment from the oxygenator. These observations were treated as checks of experimental conditions rather than as primary outcome measures.
The working-fluid volume in the VACC circuit reservoir was monitored visually using the volume graduations on the CAPIOX FX25 integrated venous reservoir. Starting from an initial volume of 100 mL, a decrease of ≥10 mL was judged to have occurred when the liquid level reached 90 mL or lower. When this criterion was met during measurement, measurements at that and subsequent higher pump flow-rate settings were not performed under the corresponding condition. This criterion represented a 10% decrease from the initial working-fluid volume in the VACC circuit reservoir. Similarly, subsequent higher pump flow-rate settings were not evaluated if a risk of air entrainment into the venous drainage line was identified or if stable maintenance of the reservoir level was judged to be difficult by visual inspection. Unmeasured data were not imputed and are presented as not determined (ND). ND values were not treated as randomly missing data. Rather, they were regarded as informative missing data reflecting difficulty in continuing measurements under the corresponding conditions with the D-pump inactive.
Statistical analysis
Statistical analyses, generation of S1 and S2 Tables, and preparation of the figures were performed using Python (version 3.12.13) in Google Colaboratory. The analyses were conducted using NumPy (version 2.0.2), pandas (version 2.2.2), SciPy (version 1.16.3), statsmodels (version 0.14.6), and Matplotlib (version 3.10.0). The versions of Python and each package were also documented in the analysis code. The analysis code is provided as S1 File.
The raw data were imported from a CSV file into Python and converted into a long-format dataset corresponding to each experimental condition, technical replicate, and measurement variable. The mean, standard deviation, standard error, and 95% confidence interval for each condition were calculated from the same processed dataset. S1 and S2 Tables were generated from the raw data using the same analysis code. Figs 2–4 and S1 and S2 Figs were generated from the same dataset and code. No manual modifications were made to numerical values, points, lines, error bars, or unmeasured conditions after the tables and figures had been generated.
Each point represents the monitor-displayed mean Pv value for a technical replicate under the corresponding experimental condition. Within each Prp,set–Prc,set combination, individual points include all technical replicates from the common pump flow-rate setting range of 50, 100, and 150 mL/min, which was available for all combinations. Diamonds indicate pooled means, and error bars indicate SDs. Pv values were stratified according to Prp,set. Pv denotes patient-side mock-reservoir outlet pressure, Prp,set denotes the patient-side mock-reservoir negative-pressure setpoint, and Prc,set denotes the circuit-reservoir negative-pressure setpoint.
Panels (a), (b), and (c) show the Prp,set conditions of −10, −5, and 0 mmHg, respectively. Points indicate means of six technical replicates, and error bars indicate SDs. Only acquired pump flow-rate settings are shown; unmeasured higher settings were not imputed. Pv denotes patient-side mock-reservoir outlet pressure, Prp,set denotes the patient-side mock-reservoir negative-pressure setpoint, and Prc,set denotes the circuit-reservoir negative-pressure setpoint. Absence of points at higher pump flow-rate settings indicates that measurements were not performed beyond the prespecified termination criteria, rather than random missing data.
Panels (a), (b), and (c) show the Prp,set conditions of −10, −5, and 0 mmHg, respectively. Points indicate means of six technical replicates, and error bars indicate SDs. Under the standardized arterial-line occluder condition, Po increased with the pump flow-rate setting. Po denotes roller pump outlet pressure, Prp,set denotes the patient-side mock-reservoir negative-pressure setpoint, and Prc,set denotes the circuit-reservoir negative-pressure setpoint. This figure shows the dependence of Po on the pump flow-rate setting under the fixed arterial-line occluder condition; it does not represent a relationship with directly measured circuit flow.
Continuous variables are presented as the mean, standard deviation, and 95% confidence interval for six technical replicates under each condition. The 95% confidence intervals were calculated using the t distribution to summarize the precision of condition-specific technical measurements within this bench system. They were not used for hypothesis testing and should not be interpreted as inferential estimates for a broader biological or clinical population.
The primary analysis was descriptive. In the exploratory analyses, an estimation-based approach was emphasized [22]. Effect sizes and 95% confidence intervals were reported. Differences were not interpreted solely on the basis of p-values. Effect sizes were defined as unstandardized effect sizes. For Pv, the effect size was defined as the mean difference between Prp,set conditions. For Po, the effect size was defined as the pressure change associated with a 100-mL/min increase in the pump flow-rate setting. For Pc, the difference between measured Pc and the corresponding Prc,set value was summarized. For Pi, changes in roller pump inlet pressure with increasing pump flow-rate settings were evaluated as a supportive analysis using S2 Fig.
For comparisons of Pv among Prp,set conditions, condition-specific means were used for the Prc,set values of −40, −30, and −20 mmHg. The pump flow-rate settings of 50, 100, and 150 mL/min were used because these combinations were available for all Prp,set conditions. Differences among conditions were calculated for corresponding combinations of Prc,set and pump flow-rate setting and are presented as mean differences with 95% confidence intervals.
The association between Po and the pump flow-rate setting was summarized using regression analyses based on condition-specific means. A pooled linear regression using all acquired condition-specific means was retained as a descriptive summary of the overall association. To assess sensitivity to unequal evaluability at higher pump flow-rate settings, a condition-adjusted linear regression was additionally fitted with the Prp,set–Prc,set combination included as a categorical fixed effect. A balanced sensitivity analysis was also restricted to pump flow-rate settings of 50, 100, and 150 mL/min. This common range was available for all 12 Prp,set–Prc,set combinations. The same condition-adjusted model was applied, and condition-specific slopes within this common range were calculated descriptively. Because visual inspection of the scatterplot suggested curvature, a regression model including a quadratic term was also evaluated as a sensitivity analysis. Unmeasured data were not imputed. Comparisons among conditions were interpreted only within the range of pump flow-rate settings for which measurements were available across the relevant conditions. Regression confidence intervals were interpreted as descriptive measures of precision within this bench dataset and not as population-inferential estimates. Full results of the regression analyses, including all estimated coefficients, standard errors, p-values, 95% confidence intervals, and goodness-of-fit measures, are provided in S2 Table.
No outliers were excluded. All acquired data were included in the analyses. The replicate-level dataset used for all analyses is provided as S1 Dataset. The condition-specific summary statistics are provided as S1 Table. The analysis notebook that reproduces the statistical analyses, generation of S1 and S2 Tables, and preparation of Figs 2–4 and S1 and S2 Figs is provided as S1 File. All supporting files are also available on Zenodo (Version v2, https://doi.org/10.5281/zenodo.22604656).
Ethics statement
This study was a bench study using a mock circulation circuit and a 32 wt% aqueous glycerol solution. No human participants, patient information, animals, animal-derived materials, or clinical specimens were used. Therefore, this study was not subject to institutional review for research involving humans or animals.
Results
Overview of acquired conditions
Before the main pressure experiments, volumetric verification showed close agreement between the pump flow-rate settings and delivered flow. Mean measured flows were 100.0 ± 0.6, 200.7 ± 0.8, and 300.8 ± 0.8 mL/min at pump flow-rate settings of 100, 200, and 300 mL/min, respectively. Each value was based on six repeated measurements. The Prp,set values examined were 0, −5, and −10 mmHg, and the Prc,set values examined were 0, −10, −20, −30, and −40 mmHg. Prp,set was treated as an operational factor, and the reduced patient-side effective filling pressure conditions were confirmed using measured Pv. The maximum pump flow-rate setting that could be evaluated differed among conditions. Some conditions were evaluated from 50 to 300 mL/min, whereas others were evaluated from 50 to 200 mL/min or from 50 to 150 mL/min. These ranges were treated as the evaluable pump flow-rate setting ranges with the D-pump inactive. The evaluable pump flow-rate setting range and measurement status for each condition are shown in Table 2. The absence of data points at higher pump flow-rate settings in several conditions does not represent randomly missing data. Rather, measurements were not performed beyond the prespecified termination criteria. Reservoir-level decrease, unstable reservoir-level maintenance, and potential air entrainment risk were assessed visually and could overlap. Therefore, the specific trigger for termination was not further subclassified.
Overall trends in Pc, Pv, Pi, and Po
Pc primarily reflected the Prc,set conditions, with mean values ranging approximately from −41.5 to 0.0 mmHg. This finding indicates that the specified Prc,set conditions were generally reproduced as circuit-reservoir pressures. Pv decreased stepwise according to Prp,set. Pv ranged approximately from −0.5 to 0.0 mmHg at Prp,set 0 mmHg, from −6.3 to −5.0 mmHg at Prp,set −5 mmHg, and from −10.0 to −9.0 mmHg at Prp,set −10 mmHg. Pi was affected by Prp,set and tended to shift in the negative direction as the pump flow-rate setting increased, ranging overall from −13.0 to 0.5 mmHg. Po consistently increased with the pump flow-rate setting and ranged overall from 2.3 to 117.0 mmHg. The mean values, standard deviations, and 95% confidence intervals for each pressure measurement site are provided in S1 Table. Table 3 summarizes the ranges of condition-specific mean Pc, Pv, Pi, and Po at representative pump flow-rate settings of 50 and 300 mL/min, stratified by Prp,set. The responses of Pc and Pi to the pump flow-rate setting are shown in S1 and S2 Figs, respectively.
Distribution of Pv and response of Pv to the pump flow-rate setting
Fig 2 shows the distribution of individual Pv values according to Prp,set within the common pump flow-rate setting range of 50, 100, and 150 mL/min. This range was available for all Prp,set–Prc,set combinations. Pv values were separated among the three Prp,set conditions of 0, −5, and −10 mmHg and decreased stepwise according to the experimental setting. Fig 3 shows the response of Pv to the pump flow-rate setting. Within each Prp,set condition, changes in Pv with increasing pump flow-rate settings were small, and differences among Prc,set conditions were limited.
Response of Po to the pump flow-rate setting
Fig 4 shows the response of Po to the pump flow-rate setting as mean ± SD. Po increased with the pump flow-rate setting under all Prp,set conditions. At a pump flow-rate setting of 50 mL/min, Po ranged from 2.3 to 10.8 mmHg. Among the conditions that could be evaluated up to 300 mL/min, Po ranged from 114.0 to 117.0 mmHg at 300 mL/min. Differences associated with Prp,set and Prc,set conditions were small compared with the changes associated with the pump flow-rate setting.
Exploratory estimation analysis
In the exploratory estimation analysis, condition-specific means calculated from six technical replicates under each condition were used as the units of analysis. Comparisons of Pv among Prp,set conditions were based on nine combinations that were evaluable across all Prp,set conditions. These combinations comprised Prc,set values of −40, −30, and −20 mmHg and pump flow-rate settings of 50, 100, and 150 mL/min.
The condition-specific mean Pv values were 0.0 mmHg at Prp,set 0 mmHg, −5.7 mmHg at Prp,set −5 mmHg, and −9.2 mmHg at Prp,set −10 mmHg. The mean difference between Prp,set −5 mmHg and Prp,set 0 mmHg was −5.7 mmHg (95% CI: −6.0 to −5.3 mmHg). The mean difference between Prp,set −10 mmHg and Prp,set 0 mmHg was −9.2 mmHg (95% CI: −9.5 to −8.9 mmHg). The mean difference between Prp,set −10 mmHg and Prp,set −5 mmHg was −3.6 mmHg (95% CI: −3.9 to −3.2 mmHg). These findings indicate that measured Pv decreased stepwise according to Prp,set.
Correspondence between selected pump flow-rate settings and delivered flow was independently verified volumetrically before the main experiments. However, delivered flow was not continuously measured during the pressure experiments. Therefore, the observed association between Po and the pump flow-rate setting was interpreted primarily as a device-setting response under the fixed occluder condition rather than as a directly measured hydraulic pressure–flow relationship. The association between Po and the pump flow-rate setting was evaluated using 64 acquired condition-specific means. This pooled regression analysis was performed to descriptively summarize the dominant association between Po and the pump flow-rate setting across acquired conditions. It was not intended to imply that Prp,set or Prc,set had no effect under all possible circuit conditions. In the linear regression analysis, Po was positively associated with the pump flow-rate setting. The regression slope was 0.430 mmHg/(mL/min) (95% CI: 0.413 to 0.447 mmHg/(mL/min)). This corresponded to an increase of 43.0 mmHg (95% CI: 41.3 to 44.7 mmHg) per 100-mL/min increase in the pump flow-rate setting. The coefficient of determination was R2 = 0.976. In the condition-adjusted analysis using all 64 acquired condition-specific means, the estimated slope was 0.428 mmHg/(mL/min) (95% CI: 0.410 to 0.446 mmHg/(mL/min)). The balanced sensitivity analysis was restricted to 50, 100, and 150 mL/min, for which all 12 Prp,set–Prc,set combinations were evaluable. The condition-adjusted slope was 0.317 mmHg/(mL/min) (95% CI: 0.304 to 0.329 mmHg/(mL/min)). The corresponding condition-specific slopes ranged from 0.302 to 0.332 mmHg/(mL/min) across the 12 Prp,set–Prc,set combinations. Thus, a positive association between pump flow-rate setting and Po was consistently observed within the common evaluable range. This finding was not driven preferentially by conditions that remained evaluable at higher pump flow-rate settings. In a sensitivity analysis restricted to conditions that could be evaluated up to 300 mL/min, the regression slope was 0.433 mmHg/(mL/min) (95% CI: 0.415 to 0.452 mmHg/(mL/min)). This estimate was similar to the pooled and all-data condition-adjusted estimates. The direction of the association was consistent across the linear analyses, although the slope magnitude differed according to the analytical range and condition structure. These regression confidence intervals should be interpreted in light of the whole-mmHg recording granularity and the technical-replicate design. They describe within-system precision and should not be interpreted as population-level uncertainty.
However, the association between Po and the pump flow-rate setting suggested curvature. In the regression model including a quadratic term, the coefficient of the quadratic term was 8.29 × 10−4 mmHg/(mL/min)2 (95% CI: 7.36 × 10−4 to 9.23 × 10−4). The adjusted R2 was 0.996. Thus, Po increased monotonically with the pump flow-rate setting, but the association was not strictly linear and included a tendency toward a greater increase in Po at higher pump flow-rate settings.
For Pc, the difference between measured Pc and the corresponding Prc,set value was evaluated across all conditions. The mean difference between Pc and Prc,set was −0.18 mmHg (95% CI: −0.37 to 0.00 mmHg), with a range from −2.0 to 1.8 mmHg. This finding indicates that the Prc,set values were generally reflected in the circuit-reservoir pressures.
Discussion
Main findings
This study descriptively characterized pressure behavior in a low-flow, roller pump-driven VACC mock circuit under standardized bench conditions with the D-pump inactive. Pc primarily reflected Prc,set. This confirmed that the experimental circuit-reservoir negative-pressure conditions were reproduced in the measured circuit-reservoir pressure. Pv served as an empirical indicator of the reduced patient-side effective filling pressure conditions created by Prp,set. Pi was interpreted as a supportive indicator of roller pump inlet-side loading. Po varied primarily according to the pump flow-rate setting under a fixed arterial-side boundary condition. These findings should be interpreted as pressure-characterization data obtained from a mock circuit and not as evidence of clinical feasibility, clinical safety, or the stable operating range of VACC.
The present findings were based on a descriptive evaluation centered on pressure data. They do not define the stable operating range of VACC. Venous drainage maintenance, reservoir-level changes, flow fluctuations, one-way valve behavior, and D-pump operating conditions were not integrated into the present analysis. These factors should be addressed in subsequent bench studies. Within this defined bench setting, the standardized manipulation of Prc,set, Prp,set, and pump flow-rate settings allowed Pc, Pv, Pi, and Po to be assigned distinct interpretive roles. Previous VACC reports primarily described the development and clinical implementation of adult centrifugal pump-driven systems [7,8], whereas prior bench work on impaired venous drainage focused on venous collapse and flow limitation [10]. In contrast, the present study separately manipulated Prc,set, Prp,set, and the pump flow-rate setting in a small-bore, roller pump-driven low-flow VACC mock circuit while measuring Pc, Pv, Pi, and Po. Thus, the new information provided here is a site-specific pressure-response framework under controlled low-flow conditions rather than evidence of clinical VACC performance. The principal contribution of this study is not the demonstration of VACC performance, but the provision of a structured framework for interpreting pressure responses in a roller pump-driven low-flow VACC mock circuit.
Interpretation of the reduced patient-side effective filling pressure condition in the bench model
The negative pressure applied to the patient-side mock reservoir was not intended to faithfully reproduce impaired venous drainage in clinical practice. Rather, it was used to simplify reductions in the driving force supplying fluid from the patient side to the venous drainage line in a manner that allowed comparisons across conditions. Venous drainage conditions during CPB may affect arterial flow and circuit pressure. Negative-pressure conditions require careful management of reservoir pressure, air entrainment, microbubbles, and hemolysis [11–13]. Impaired venous drainage and venous collapse have also been shown in a bench model to contribute to flow limitation during CPB [10]. However, pressure at the patient-side mock-reservoir outlet was affected by the reservoir fluid level, the measurement reference point, and hydrostatic pressure in addition to Prp,set. Therefore, Prp,set itself cannot be interpreted as the actual pressure within the venous drainage line. Measured Pv was used as the empirical indicator of the imposed patient-side condition.
Manipulation of patient-side mock-reservoir pressure resulted in a stepwise decrease in Pv. This finding indicates that measured Pv served as an empirical indicator of reduced patient-side effective filling pressure conditions within this model. Accordingly, patient-side conditions were interpreted on the basis of measured Pv rather than the patient-side mock-reservoir pressure setpoint itself.
This model was based on static reservoir conditions. It did not incorporate venous capacitance, vascular collapse, body position, cannula position, or changes in blood properties in actual small dogs. Therefore, the findings do not directly predict clinical conditions and should be interpreted as fundamental data for understanding pressure behavior in low-flow VACC circuits.
Interpretation of pressure behavior in the low-flow range
In this study, Pc, Pv, Pi, and Po were positioned as pressure indicators that primarily reflected different experimental conditions. Pc served as an internal validity indicator confirming the reflection of Prc,set conditions within the experimental setup. Pv was stratified according to Prp,set. Condition-specific means within the common evaluation range showed clear mean differences among Prp,set conditions. Therefore, measured Pv, rather than the patient-side mock-reservoir pressure setpoint itself, can be interpreted as an indicator of reduced patient-side effective filling pressure conditions in this model.
Pi represents roller pump inlet pressure and may be affected by patient-side effective filling pressure, the pump flow-rate setting, inlet-side circuit resistance, and the degree of roller pump occlusion. The degree of tubing occlusion in a roller pump is related to the residual flow-channel geometry within the compressed tubing and to mechanical stress [23]. In extracorporeal circulation, insufficient occlusion may result in backflow or reduced flow, whereas excessive occlusion may increase tubing load and mechanical stress. Contemporary CPB guidelines likewise emphasize that roller-pump performance depends on appropriate occlusion adjustment [24]. They also note that underocclusion may result in inaccurate flow readout, reinforcing the distinction between pump-indicated flow and independently verified delivered flow [24]. The same guidelines emphasize appropriate calibration and monitoring of arterial pump flow and pressure transducers [24]. This further supports a careful distinction between device settings, measured flow, and measured pressure when interpreting extracorporeal-circuit data. The importance of quantitatively evaluating and documenting roller pump occlusion has been demonstrated in both the original FIP study and a subsequent evaluation using clinical CPB equipment [20,25]. In the present study, the degree of occlusion was adjusted before testing and remained unchanged to prevent occlusion from acting as a confounding factor when comparing pressure behavior across conditions.
Pi tended to shift in the negative direction as the pump flow-rate setting increased. This finding suggests that changes in roller pump inlet-side load associated with increasing pump flow-rate settings may be reflected in Pi, even within the low-flow range. However, because Pi is affected by multiple factors, it should not be interpreted as a standalone indicator of venous drainage stability. Future evaluations should interpret Pi together with Pv, venous drainage flow, flow fluctuations, and reservoir-level changes.
Po consistently increased with the pump flow-rate setting. Under the fixed arterial-line occluder condition, differences associated with Prp,set and Prc,set were relatively small. The exploratory regression analysis showed a high coefficient of determination for the association between Po and the pump flow-rate setting. However, the absolute value of Po may be affected by the measurement location, zeroing position, vertical distance from the reservoir fluid level, oxygenator, and occluder setting. Therefore, the present findings should not be interpreted as clinical safety thresholds. Although correspondence between selected pump flow-rate settings and delivered flow was verified before the main experiments, delivered flow was not continuously measured during the pressure experiments. The upward curvature observed in the Po response may reflect an increasing contribution of flow-dependent hydraulic losses across the fixed arterial-line occluder, tubing, connectors, and other circuit components at higher pump flow-rate settings. Pressure-dependent deformation of the tubing may also have contributed. However, because delivered flow was not continuously measured during the pressure experiments, the relative contributions of these mechanisms could not be separated in the present study. The quadratic model was used only as an exploratory sensitivity analysis and was not intended to establish a specific hydraulic mechanism. Accordingly, Po should be interpreted as roller pump outlet pressure relative to the reference height defined in this experimental setup, not as actual arterial-line pressure or as a hydraulic pressure–flow relationship.
Interpretation of pressure behavior with the D-pump inactive
Although the D-pump and bubble trap were incorporated as components of the VACC circuit configuration, the D-pump remained inactive during the primary analysis. This was because D-pump operation represents an independent intervention that may affect reservoir level, circuit flow balance, and pressure behavior at all measurement sites. Accordingly, the data obtained in this study should be interpreted as observations of pressure behavior in the absence of D-pump-mediated de-airing and associated fluid return.
When the D-pump is operated, blood or circuit fluid may return to the circuit reservoir as a result of D-pump-mediated de-airing. At the same time, patient-side volume, venous drainage-side pressure, one-way valve opening and closing, and reservoir level may change in an interrelated manner. Because the D-pump remained inactive and the setup did not incorporate biological circulation, such dynamic behavior during D-pump operation was outside the scope of this study.
The present study prioritized the characterization of pressure behavior and the limits of reservoir-level maintenance with the D-pump inactive. This was intended as a preliminary step before evaluating continuous de-airing conditions with the D-pump active. Under this framework, a decrease in the reservoir level should not be regarded merely as an experimental failure. Rather, it indicates that reservoir-level maintenance was difficult under the corresponding condition with the D-pump inactive.
A decrease of ≥10 mL from the initial working-fluid volume of 100 mL in the VACC circuit reservoir was defined as a prespecified measurement termination criterion. This 10% decrease was used to apply a consistent criterion for identifying difficulty in maintaining the reservoir level with the D-pump inactive. Thus, termination of measurements under some conditions should be interpreted as indicating difficulty in continuing the present bench measurements with the D-pump inactive. It should not be interpreted as a direct indicator of clinical VACC failure or instability. The present findings should be regarded as a baseline reference condition against which future experiments with stepwise D-pump activation can be compared.
Limitations
This study has several limitations.
First, the present analysis was centered on pressure data. Correspondence between the pump flow-rate setting and delivered flow was independently verified volumetrically before the main experiments at 100, 200, and 300 mL/min. However, delivered flow was not continuously measured during the pressure experiments. The remaining settings of 50, 150, and 250 mL/min were not independently calibrated. Therefore, the study could not evaluate venous drainage adequacy, flow maintenance, flow fluctuations, or directly measured pressure–flow relationships. The pump flow-rate setting was therefore retained as the operational flow variable in the primary analysis. Potential differences between the pump flow-rate setting and delivered flow during the pressure experiments were not directly quantified, particularly at the uncalibrated settings and under varying circuit conditions. The FIP pressure of 53 mmHg at 300 s quantitatively documented the occlusion setting used in the present experiment. However, it should not be interpreted as a validated optimal threshold for the specific combination of 1/4-inch tubing and an 85-mm-diameter pump head.
Second, the maximum evaluable pump flow-rate setting differed among conditions, and not all conditions could be compared up to 300 mL/min. Therefore, comparisons among conditions involving higher pump flow-rate settings should be interpreted only within the range of pump flow-rate settings available across the relevant conditions.
Third, time-series waveforms were not analyzed. Therefore, roller pump-induced pulsatility, instantaneous pressure fluctuations, and frequency components could not be evaluated. The primary analysis used mean pressure values displayed on the physiological monitor, and these displayed means could not be recalculated from externally recorded waveforms. In addition, pressure values were recorded in whole-mmHg increments, and the technical replicates were obtained using the same bench circuit. Therefore, the narrow confidence intervals observed in some analyses should be interpreted as reflecting measurement granularity and repeated-reading consistency within this experimental system rather than uncertainty in an underlying biological or population parameter.
Fourth, the working fluid was an aqueous glycerol solution. Aqueous glycerol solutions can be used as mock fluids approximating blood viscosity [15]. However, they do not reproduce the non-Newtonian properties of blood, cellular components, hemolysis, coagulation, or vascular compliance [16,18].
Fifth, although the order of Prp,set and Prc,set condition combinations was randomized, pump flow-rate settings were evaluated in ascending order within each condition combination. Because the pump flow-rate settings were always evaluated in ascending order within each replicate run, the effect of pump flow-rate setting could not be completely separated from elapsed time, cumulative tubing conditioning, or other order-dependent changes within a run. Reservoir-level changes, potential air entrainment risk, and obvious backflow were assessed visually by two experimenters but were not evaluated as blinded primary outcome measures. Therefore, the termination status and visual observations should be interpreted as experimental condition checks within the present bench setup. Observer-dependent judgment cannot be completely excluded. These observations were not used to establish safety thresholds, stable operating limits, or clinical failure criteria. In addition, although the patient-side mock-reservoir outlet was fixed 10 cm above the common pressure-reference height, the instantaneous fluid-surface height was not recorded separately. Therefore, the hydrostatic contribution to measured Pv could not be quantitatively separated from Prp,set and other circuit effects.
Accordingly, the findings of this study should be interpreted as fundamental mock-circuit pressure-characterization data and should not be used to infer the stable operating range, clinical safety, clinical efficacy, or generalized performance of VACC.
Future directions
The next step should remain within the bench-testing framework. Future bench studies should prespecify and integrate directly measured flow, criteria for the maximum maintainable pump flow-rate setting, reservoir-level changes, and flow variation or pulsatility indices. Reservoir-return flow, one-way valve behavior, and D-pump operating status should also be prespecified and integrated. Such data would allow the descriptive pressure-characterization data presented in this study to be extended to a quantitative mock-circuit evaluation.
Further bench studies incorporating stepwise D-pump operation are warranted. These studies should clarify how the limits of reservoir-level maintenance and pressure behavior observed with the D-pump inactive change when D-pump-mediated de-airing and associated fluid return are included. In future estimation-based analyses, estimated differences, 95% confidence intervals, and effect sizes should be reported. These measures would allow the magnitude and uncertainty of differences among conditions to be evaluated.
The present study characterized pressure behavior within a low-flow range of 50–300 mL/min. In CPB for low-body-weight patients, priming volume and circuit miniaturization have also been examined as important technical issues in human pediatric and neonatal practice [26]. However, this study did not evaluate pediatric or adult human CPB circuits. The working fluid, circuit dimensions, flow-rate range, and pressure boundary conditions differ across these contexts. Therefore, the present findings cannot be directly extrapolated to other fields. Future studies should clearly define the target circuit and pressure boundary conditions when extending pressure-characterization data to other low-flow mock-circuit settings.
Conclusions
In a low-flow, roller pump-driven VACC mock circuit with the D-pump inactive, Pc reflected the circuit-reservoir negative-pressure setpoint. Measured Pv served as an empirical indicator of experimentally reduced patient-side effective filling pressure conditions. Pi shifted in the negative direction with increasing pump flow-rate settings. It should be interpreted only as a supportive indicator of roller pump inlet-side loading. Po increased according to the pump flow-rate setting under a fixed arterial-side boundary condition. Although correspondence between selected pump flow-rate settings and delivered flow was independently verified before the main experiments, delivered flow was not continuously measured during the pressure experiments. Therefore, Po was interpreted primarily in relation to the pump flow-rate setting. Together, these findings provide a practical pressure-characterization framework for future bench studies incorporating directly measured flow, reservoir-level dynamics, one-way valve behavior, and D-pump-assisted operation. They do not establish the stable operating range, clinical safety, clinical efficacy, or generalized performance of VACC.
Supporting information
S1 Dataset. Replicate-level dataset. Raw data used for all analyses, including replicate identifiers, pump flow-rate settings, Prp,set, Prc,set, Pc, Pv, Pi, and Po for each acquired experimental condition.
https://doi.org/10.1371/journal.pone.0363658.s001
(CSV)
S1 Table. Summary statistics for Pc, Pv, Pi, and Po according to Prp,set, Prc,set, and pump flow-rate setting conditions.
Mean, standard deviation, n, standard error, and 95% confidence interval are presented for each condition.
https://doi.org/10.1371/journal.pone.0363658.s002
(DOCX)
S2 Table. Full results of regression analyses for Po.
Estimated regression coefficients, standard errors, p-values, 95% confidence intervals, and goodness-of-fit measures are presented for the pooled linear regression, the condition-adjusted linear regression, the balanced common-range condition-adjusted sensitivity analysis, the sensitivity analysis restricted to condition combinations evaluated through 300 mL/min, and the quadratic regression sensitivity analysis.
https://doi.org/10.1371/journal.pone.0363658.s003
(DOCX)
S1 File. Analysis notebook.
Google Colab-compatible Jupyter notebook reproducing the statistical analyses, generation of S1 and S2 Tables, and preparation of Figs 2–4 and S1 and S2 Figs.
https://doi.org/10.1371/journal.pone.0363658.s004
(IPYNB)
S1 Fig. Response of Pc to the pump flow-rate setting according to Prp,set conditions.
Pc is presented as mean ± SD according to the pump flow-rate setting under each combination of Prp,set and Prc,set. Only acquired pump flow-rate settings are shown; unmeasured conditions were not imputed. Pc denotes circuit-reservoir pressure, Prp,set denotes the patient-side mock-reservoir negative-pressure setpoint, and Prc,set denotes the circuit-reservoir negative-pressure setpoint. This figure is intended to show whether Pc reflected the Prc,set conditions across the acquired pump flow-rate settings.
https://doi.org/10.1371/journal.pone.0363658.s005
(TIF)
S2 Fig. Response of Pi to the pump flow-rate setting according to Prp,set conditions.
Pi is presented as mean ± SD according to the pump flow-rate setting under each combination of Prp,set and Prc,set. Only acquired pump flow-rate settings are shown; unmeasured conditions were not imputed. Pi denotes roller pump inlet pressure, Prp,set denotes the patient-side mock-reservoir negative-pressure setpoint, and Prc,set denotes the circuit-reservoir negative-pressure setpoint. Pi was interpreted as a supportive indicator of roller pump inlet-side loading and was not used to define drainage failure or a stable operating range.
https://doi.org/10.1371/journal.pone.0363658.s006
(TIF)
Acknowledgments
The authors thank Chisato Ikubo of the Graduate Degree Program of Comprehensive Applied Data Science, Teikyo University, and the undergraduate members of the Veterinary Extracorporeal Circulation & Blood Purification Lab, Clinical Engineering Course, Department of Medical Technology, Faculty of Fukuoka Medical Technology, Teikyo University, for assistance with circuit preparation, experimental setup, experimental observations, and data collection.
References
- 1. Kanemoto I, Taguchi D, Yokoyama S, Mizuno M, Suzuki H, Kanamoto T. Open heart surgery with deep hypothermia and cardiopulmonary bypass in small and toy dogs. Vet Surg. 2010;39(6):674–9. pmid:20459489
- 2. Uechi M, Mizukoshi T, Mizuno T, Mizuno M, Harada K, Ebisawa T, et al. Mitral valve repair under cardiopulmonary bypass in small-breed dogs: 48 cases (2006-2009). J Am Vet Med Assoc. 2012;240(10):1194–201. pmid:22559109
- 3. Wauchi K, Yoshitake S, Hiroura M, Yoshikae R, Sakai K, Fujioka T. Substitution of dialysis needle catheter for blood-feeding catheter in artificial heart-lung in a toy breed dog. Adv Anim Cardiol. 2023;56:75–83.
- 4. Kanemoto I, Mihara K, Sato K. Open-heart techniques and mitral valve plasty for mitral regurgitation in toy- and small-breed dogs: A review. Open Vet J. 2021;11(1):14–26. pmid:33898279
- 5. Pelosi A, Anderson LK, Paugh J, Robinson S, Eyster GE. Challenges of cardiopulmonary bypass-a review of the veterinary literature. Vet Surg. 2013;42(2):119–36. pmid:23164065
- 6. Gajkowski EF, Herrera G, Hatton L, Velia Antonini M, Vercaemst L, Cooley E. ELSO guidelines for adult and pediatric extracorporeal membrane oxygenation circuits. ASAIO J. 2022;68(2):133–52. pmid:35089258
- 7. Okumura T, Matsuda K, Fukuoka Y, Dai J, Shiraishi N. Development of a valve type semi-closed extracorporeal circulation system. J Artif Organs. 2021;24(3):320–6. pmid:33534081
- 8. Matsuda K, Okumura T, Chiba H, Ujiie A, Kosawa Y, Kamezawa S. Introduction of the valve type semi-closed extracorporeal circulation method: Seeking an experience-independent cardiopulmonary bypass. Jpn J Extra-Corporeal Technology. 2022;49:117–23.
- 9. Aizawa K, Negishi M, Machiyama A, Okumura T. Study of accidental decannulation of an air lock in valve-type, semi-closed, extracorporeal circulation. Transactions Japanese Society for Med Biological Eng. 2022;60:164–9.
- 10. Li L, Abdel-Sayed S, Berdajs D, Tozzi P, von Segesser LK, Ferrari E. Caval collapse during cardiopulmonary bypass: A reproducible bench model. Eur J Cardiothorac Surg. 2014;46(2):306–12. pmid:24431162
- 11. Carvalho Filho EB de, Marson FA de L, Costa LNG da, Antunes N. Vacuum-assisted drainage in cardiopulmonary bypass: Advantages and disadvantages. Rev Bras Cir Cardiovasc. 2014;29(2):266–71. pmid:25140478
- 12. Gambino R, Searles B, Darling EM. Vacuum-assisted venous drainage: A 2014 safety survey. J Extra Corpor Technol. 2015;47(3):160–6. pmid:26543250
- 13. Jones TJ, Deal DD, Vernon JC, Blackburn N, Stump DA. Does vacuum-assisted venous drainage increase gaseous microemboli during cardiopulmonary bypass?. Ann Thorac Surg. 2002;74(6):2132–7. pmid:12643407
- 14.
Nihon Kohden Corporation. BSM-2300 series bedside monitor service manual. 9th ed. Tokyo: Nihon Kohden Corporation; 2014.
- 15. Boës S, Ochsner G, Amacher R, Petrou A, Meboldt M, Schmid Daners M. Control of the Fluid Viscosity in a Mock Circulation. Artif Organs. 2018;42(1):68–77. pmid:28718516
- 16. Rand PW, Lacombe E. Hemodilution, tonicity, and blood viscosity. J Clin Invest. 1964;43:2214–26.
- 17. Volk A, Kähler CJ. Density model for aqueous glycerol solutions. Exp Fluids. 2018;59(5).
- 18. Popel AS, Johnson PC. Microcirculation and Hemorheology. Annu Rev Fluid Mech. 2005;37:43–69. pmid:21151769
- 19.
Japanese Standards Association. JIS T 1603:1995. Electric motor driven blood pump for cardiopulmonary bypass. Tokyo: Japanese Standards Association. 1995.
- 20. Okumura T, Nakazawa Y, Machida T. Quantification of the degree of occlusion of roller pumps by the fixed flow rate-injection pressure method: Basic research for quantifying the degree of occlusion. Jpn J Extra-Corporeal Technology. 2021;48:62–7.
- 21. Vaux DL, Fidler F, Cumming G. Replicates and repeats—what is the difference and is it significant?. EMBO Rep. 2012;13(4):291–6.
- 22. Ho J, Tumkaya T, Aryal S, Choi H, Claridge-Chang A. Moving beyond P values: Data analysis with estimation graphics. Nat Methods. 2019;16(7):565–6. pmid:31217592
- 23. Fukaya A, Shiraishi Y, Inoue Y, Yamada A, Sahara G, Kudo T, et al. Development and accuracy evaluation of a degree of occlusion visualization system for roller pumps used in cardiopulmonary bypass. J Artif Organs. 2021;24(1):27–35. pmid:32930908
- 24. Wahba A, Kunst G, De Somer F, Agerup Kildahl H, Milne B, Kjellberg G, et al. 2024 EACTS/EACTAIC/EBCP Guidelines on cardiopulmonary bypass in adult cardiac surgery. Eur J Cardiothorac Surg. 2025;67(2):ezae354. pmid:39949326
- 25. Kodama K, Aizawa K, Inoue M, Shiraishi N, Okumura T. Quantitative evaluation of appropriate roller-pump occlusion adjusted by conventional methods using the fixed flow rate-injection pressure method: verification using clinical equipment [in Japanese]. J Japan Association for Clinical Engineers. 2023;80:72–6.
- 26. Miyaji K, Miyamoto T, Kohira S, Nakashima K, Inoue N, Sato H, et al. Miniaturized cardiopulmonary bypass system in neonates and small infants. Interact Cardiovasc Thorac Surg. 2008;7(1):75–8. pmid:18003721