Three-Dimensional Ultrasound Localization Microscopy Using Hadamard-Encoded Multipulse and 4D Denoising on a 256-Channel System with a 1024-element Matrix Array: A Feasibility Study in Human Kidney Transplants
Abstract
Three-dimensional ultrasound localization microscopy (3D ULM) enables volumetric visualization of microvasculature beyond the acoustic diffraction limit, but its clinical translation remains limited by long acquisition times, low microbubble (MB) signal-to-noise ratio (SNR), and insufficient MB contrast, particularly on ultrasound systems with a limited number of acquisition channels.
In this study, we developed an in-human 3D ULM framework for a 256-channel research ultrasound system equipped with a 1024-element matrix array.
The framework combines diverging-wave Hadamard-encoded multipulse (HEM) transmission to enhance MB SNR with denoising strategy combining MB enhancement and four-dimensional non-local means filtering to improve MB contrast.
The framework was implemented on a research ultrasound platform and evaluated in phantom experiments and four in vivo human kidney transplants.
In phantom experiments, HEM improved the MB contrast ratio around 2 dB, while the proposed denoising strategy enhanced in vivo MB contrast by up to around 9 dB.
3D ULM reconstruction achieved a spatial resolution of approximately 84 um, as estimated by Fourier shell correlation.
The number of reconstructed microvascular voxels increased exponentially with acquisition duration, reaching approximately 74% of the estimated saturated vessel density using 3000 volumes (~12 s).
The subject with the highest histopathological injury exhibited quantitatively reduced microvascular vessel density, flow velocities, and mean perfusion index.
These results demonstrate that the proposed framework enables high-quality volumetric super-resolution imaging and quantitative characterization of renal microvasculature using a 256-channel ultrasound system, providing a practical step toward the clinical translation of in vivo 3D ULM.