Single-Pulse Dosimetry for Ultra-High-Dose-Rate X-Ray Pulses Based on Radiation-Induced Acoustic Signals
Abstract
Real-time dosimetry of ultra-high-dose-rate (UHDR) pulsed X-rays remains a major challenge.
This proof-of-principle study investigates a dosimetry method for UHDR pulsed X-rays based on radiation-induced acoustic (RA) signals from a thin metal foil.
The foil combines efficient X-ray absorption and acoustic resonance, converting deposited dose into acoustic signals with discrete resonant frequencies.
Single-pulse measurements were performed at six dose levels spanning approximately 1--40 mGy(Si), with one pulse recorded at each level and instantaneous dose rates reaching the order of 10^6 Gy/s at the highest dose settings.
The peak-to-peak amplified transducer-output voltage exhibited a linear relationship with the reference dose measured by thermoluminescent dosimeters (TLDs).
These results demonstrate the feasibility of single-pulse acoustic dosimetry under the tested conditions.
Furthermore, we establish analytical solutions for radiation-induced acoustic signals in a one-dimensional layered medium in both the time and frequency domains, providing a theoretical basis for signal interpretation and detector optimization.