Patient acceptance of continuous wearable vital-sign monitoring on surgical wards: A prospective mixed-methods survey study
Figures
Abstract
Background
Continuous wearable monitoring of vital signs is increasingly implemented on hospital wards to enable early detection of clinical deterioration and to reduce the burden of intermittent vital-sign measurements. While feasibility of wearable monitoring and its acceptance among healthcare professionals have been demonstrated, patient perspectives remain underexplored.
Methods
This prospective, single-center mixed-methods study included adult surgical ward patients monitored with a wearable device (Healthdot or viQtor) for ≥ 24 hours. Patients completed 15 five-point Likert items assessing patient experience and a separate overall satisfaction rating on a 1–10 scale. Qualitative material comprised free-text responses from 36 of the 102 participants and two additional semistructured interviews with participants from the same cohort. Quantitative data were analyzed using descriptive statistics, reliability analysis, exploratory monitoring-context comparisons, and multivariable regression. Qualitative data were analyzed using Braun and Clarke’s thematic-analysis approach.
Results
A total of 102 patients were included in the final analysis after exclusion of two duplicate questionnaire records. Mean overall acceptance was 4.05 ± 0.54 on a 1–5 scale (95% CI 3.94–4.15), and mean satisfaction was 8.21 ± 1.39 on a 1–10 scale (95% CI 7.93–8.48). Acceptance was higher in the Healthdot context than in the viQtor context, with an exploratory mean difference of 0.40 points (95% CI 0.18–0.62). Qualitative findings indicated generally limited burden, but also concerns regarding comfort and patient information.
Conclusions
Patients in this single-center cohort generally reported favorable experiences with continuous wearable vital-sign monitoring. Acceptance differed between the two monitoring contexts, but the study design does not allow these differences to be attributed to a causal effect of device type. Further studies should validate the questionnaire and evaluate clinical effectiveness, safety, cost-effectiveness, and longer-term implementation outcomes.
Citation: Jerry EE, Nabben DLJ, De Bie Dekker AJR, Paulussen IWF, Bouwman RA, Nienhuijs SW (2026) Patient acceptance of continuous wearable vital-sign monitoring on surgical wards: A prospective mixed-methods survey study. PLoS One 21(11): e0363253. https://doi.org/10.1371/journal.pone.0363253
Editor: Gaurav Arora, King Mongkut’s University of Technology North Bangkok, THAILAND
Received: May 7, 2026; Accepted: November 5, 2026; Published: November 23, 2026
Copyright: © 2026 Jerry et al. This is an open access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
Data Availability: All relevant data are within the paper and its Supporting information files.
Funding: The author(s) received no specific funding for this work.
Competing interests: The authors have declared that no competing interests exist.
Introduction
Continuous wearable monitoring of hospitalized patients has gained increasing attention as healthcare systems aim to improve early detection of clinical deterioration while optimizing workflow efficiency. On most surgical wards, patient monitoring still relies on intermittent measurements of vital signs, typically performed every four to eight hours. Although this approach is deeply embedded in routine clinical practice, it creates temporal gaps during which early signs of physiological deterioration may remain undetected [1,2]. Delayed recognition has been associated with increased morbidity, unplanned intensive care admissions, and preventable adverse events [3].
Wearable monitoring technologies offer a promising solution to these limitations. Depending on the device, wearable systems can continuously or periodically monitor physiological parameters such as heart rate, respiratory rate, oxygen saturation, skin temperature, posture, and physical activity [4–7].
Previous studies have demonstrated the feasibility and safety of wearable monitoring in hospitalized patients, as well as its potential to support earlier detection of deterioration, improve clinical decision-making, and reduce nursing workload [6,8–12]. However, successful implementation depends not only on technological performance but also on acceptance by both healthcare professionals and patients.
Although healthcare professional perspectives have been examined in previous implementation studies, patient perspectives remain less well characterized. In the REQUEST implementation study at the same institution, manual spot checks decreased by 62.7%, corresponding to an estimated time saving of 10.1 minutes per patient per day. Mean Integrated Workload Scale scores were 5.46 at baseline, reached their lowest value of 3.87 during implementation, and were 4.62 at the final assessment. Mean System Usability Scale scores increased from 74.2 to 86.0 across assessments [9]. As primary users during hospitalization, patients’ experiences may strongly influence adherence, acceptance, and long-term sustainability of wearable monitoring [9,12]. Potential concerns include discomfort, interference with daily activities or sleep, and data privacy, whereas potential benefits include reassurance and perceived sense of safety [9,12].
Patient experience is therefore a critical factor in successful integration of wearable monitoring into clinical care. Positive experiences may facilitate broader adoption of digital health technologies, whereas negative experiences may limit their effectiveness.
The PEX-WEAR study was conducted to evaluate patient experiences with wearable monitoring during hospitalization. The main descriptive outcomes were overall acceptance and overall satisfaction. The study also explored domain-specific experiences related to usefulness, comfort, perceived safety, reassurance, sleep quality, mobility, and willingness to reuse the technology. Free-text responses and two semistructured interviews were used to provide contextual explanations for these quantitative findings. Comparisons between monitoring contexts were considered exploratory because device and ward were almost completely linked.
Materials and methods
Study design
This was a prospective, observational, single-center mixed-methods study. The study combined a quantitative patient questionnaire with free-text responses and two additional semistructured interviews to evaluate patient experiences with wearable monitoring during hospitalization.The study was reviewed by the Medical Ethics Committee United (Nieuwegein, the Netherlands), which determined that the study does not fall under the scope of the Medical Research Involving Human Subjects Act (WMO) (reference number W25.267).
Study setting and population
The study was conducted at Catharina Hospital Eindhoven, the Netherlands, on two surgical wards where wearable monitoring had been implemented as part of routine care: a general surgery ward using viQtor and an oncological surgery ward using Healthdot. The wards admitted patients from several surgical specialties, including general surgery, urology, gynecology, and oncological surgery. Each participating ward had approximately 20 beds. On both wards, the usual nurse-to-patient ratios were approximately 1:4 during the day, 1:6 during the evening, and 1:10 at night.Monitoring context and ward were strongly linked: 40 participants using viQtor were treated on the general surgery ward, whereas 61 of 62 participants using Healthdot were treated on the oncological surgery ward. Both wards were staffed by a single nursing team under one team leader. Nurses generally worked on one of the two wards but could work across both.
Device characteristics
Healthdot was a CE-marked adhesive wearable data logger placed over the left lower rib. Accelerometer-derived signals were used to estimate heart rate, respiratory rate, posture and physical activity every five minutes. Measurements were displayed as trends in the Guardian dashboard. The battery life was approximately 14 days, after which the device was replaced if necessary. viQtor was a CE-marked Class IIa remote patient-monitoring solution worn on the upper arm with an adjustable armband. The solution comprised the wearable device, charger, mobile application, web portal, and backend. It measured oxygen saturation, pulse rate, respiratory rate, skin temperature, and activity and transferred data every five minutes. The device could also detect possible falls and allow user-initiated assistance requests; these functions were not used or evaluated in this study.
Data collection
Eligible patients were identified through weekly screening of ward admission lists and expected discharge dates. Patients were approached after at least 24 hours of wearable monitoring during their hospital admission. Data collection was conducted between 20 January and 1 April 2026.
The target sample size of approximately 100 questionnaire respondents was pragmatic and intended to provide a descriptive estimate of patient acceptance. No formal target number of interviews or saturation criterion was prespecified. In total, 104 questionnaire records were screened. Records PEX030 and PEX087 were identified as duplicate records and were excluded from the analytic cohort, resulting in 102 participants for the final analysis.
Quantitative questionnaire
Patient experience was assessed using a structured questionnaire administered at the bedside during hospitalization or shortly before discharge (see S1 Table).
The questionnaire consisted of 15 five-point Likert items and a separate overall satisfaction rating on a 1–10 scale. The Likert items addressed perceived usefulness, comfort, perceived safety, reassurance, sleep quality, mobility, and willingness to reuse the technology. The questionnaire was adapted from the Technology Acceptance Model 2 [15] by a multidisciplinary panel comprising a surgeon, an anesthesiologist, an experienced researcher, and two PhD researchers. The panel selected and reviewed the items; patients were not involved in item selection. The questionnaire was not formally piloted or pretested before the study. Although TAM2 is a validated theoretical framework, the adapted questionnaire and derived acceptance score have not been formally validated in this clinical context. Construct validity was not assessed. Fourteen Likert items contributed to six theory-informed domains: perceived usefulness, comfort, perceived safety, reassurance, sleep quality, and willingness to reuse. Negatively phrased items were reverse-coded. The overall acceptance score was calculated as the unweighted mean of the six domain scores. The mobility item was analyzed separately. The separate overall satisfaction item was not included in the acceptance score.
Additional variables
Demographic and clinical data were retrieved from the electronic health record, including age, sex, BMI, surgery type, hospital stay length, and any complications during hospitalization. Device variables such as monitoring duration and wear time were also documented.
Statistical analysis
Descriptive statistics were used to summarize patient characteristics and questionnaire responses. Individual Likert items were treated as ordinal variables and summarized using response distributions, medians, and interquartile ranges. Multi-item domain scores and the overall acceptance score were summarized using means, standard deviations, and 95% confidence intervals. Internal consistency was assessed using Cronbach’s alpha.
Exploratory comparisons between Healthdot and viQtor were reported as comparisons between monitoring contexts because device and ward were almost completely linked. Exploratory between-context comparisons were performed using Welch’s t tests. Mean differences with 95% confidence intervals and Hedges’ g were reported, and Mann–Whitney U tests were used as nonparametric sensitivity analyses. Domain scores were calculated from the available items; no value was imputed for the single missing Q10 response.
A multivariable linear regression model estimated the association between monitoring context and overall acceptance, adjusting for age, sex, BMI, length of stay, and complications. Because monitoring context and ward were strongly linked, this estimate was interpreted as an adjusted association rather than a causal device effect. All 102 participants were included in the final model.
All statistical analyses were performed using IBM SPSS Statistics version 29 (IBM Corp., Armonk, NY, USA). A two-sided p-value < 0.05 was considered statistically significant.
Qualitative data collection
Qualitative material was obtained from two sources. First, 36 of the 102 questionnaire respondents provided free-text feedback. Second, two additional semistructured interviews were conducted with participants from the same 102-patient cohort. The interviews followed the questionnaire items and used open “why” prompts to elicit explanations.
The interviews were conducted in the patients’ rooms by independent physician-researchers who had no treatment relationship with the participants. Each interview lasted approximately 30 minutes. The interviews were not audio-recorded. The interviewers documented the responses as contemporaneous written notes, and no verbatim transcripts were produced.
Qualitative analysis
The qualitative material was analyzed using Braun and Clarke’s thematic-analysis approach [16]. One physician-researcher coded the material in Excel, after which the emerging themes were reviewed and discussed by the research team. No formal codebook was developed and no inter-coder reliability statistic was calculated. Differences in interpretation were resolved through discussion. Multiple categories could be assigned to each free-text response. Frequencies were calculated using the 36 questionnaire respondents who provided free-text feedback as the denominator. The interview excerpts were used to provide contextual information and were not included in these frequency counts.
Results
Study population
A total of 104 questionnaire records were screened. PEX030 and PEX087 were excluded because they represented duplicate records, resulting in 102 participants in the final analytic cohort. Of these, 62 (60.8%) used the Healthdot device and 40 (39.2%) used the viQtor device. One response was missing for Q10; all other questionnaire items were complete. All analyses were performed using the verified dataset. Baseline characteristics of the study population are summarized in Table 1.
The mean age was 65.5 ± 14.5 years, and 61 patients (59.8%) were male. The mean BMI was 26.1 ± 5.0 kg/m². The median length of hospital stay was 7 days (IQR 5–11). Patients underwent a wide range of surgical procedures, including major abdominal oncological surgery, colorectal surgery, urological surgery, and gynecological surgery. A subset of patients was admitted for non-surgical management, including treatment of infections or postoperative complications.
Main descriptive outcomes
The mean overall acceptance score was 4.05 ± 0.54 (95% CI 3.94–4.15) on a 1–5 scale. The mean overall satisfaction score was 8.21 ± 1.39 (95% CI 7.93–8.48) on a 1–10 scale. Domain means ranged from 3.50 for reassurance to 4.42 for comfort (Table 2). Cronbach’s alpha was 0.84 for the 14 scored questionnaire items (complete-item n = 101) and 0.80 for the six domain scores (n = 102).
Exploratory monitoring-context comparisons Overall acceptance was higher in the Healthdot context than in the viQtor context (mean difference 0.40, 95% CI 0.18–0.62; Hedges’ g = 0.79; p < 0.001). The largest between-context difference was observed for comfort (mean difference 0.83, 95% CI 0.47–1.19; Hedges’ g = 1.09; p < 0.001). No statistically significant between-context differences were observed for perceived safety or the single mobility item. Nonparametric sensitivity analyses showed the same overall pattern (S2 Table).
In the exploratory multivariable model, Healthdot context was associated with higher overall acceptance (adjusted regression coefficient 0.39, 95% CI 0.18–0.59; p < 0.001). The model explained 16.7% of the variance (adjusted R² = 11.4%). Age, sex, BMI, length of stay, and complications were not statistically significantly associated with overall acceptance (all p > 0.05; S2 Table).
Qualitative findings
Analysis of questionnaire free-text responses identified recurring descriptive categories related to patient experiences with wearable monitoring. Among the 36 questionnaire respondents who provided free-text feedback, experiences ranged from minimal awareness and no perceived burden to tightness, pressure, skin irritation, or device displacement. Several respondents reported limited information about the purpose of the device and expressed an interest in receiving feedback on measured values. The two semistructured interviews provided additional explanations concerning comfort, information, perceived usefulness, and reassurance. Descriptive categories and illustrative excerpts from questionnaire free-text responses are presented in Table 3.
Discussion
Patient acceptance was generally favorable in this cohort. However, patient acceptance alone does not establish clinical feasibility, effectiveness, or safety. In addition, the identified gap in patient information highlights the need for structured education and clear communication regarding the purpose and function of wearable monitoring.
A key strength of this study is its real-world design, as monitoring was already implemented as part of routine clinical care. This allows the findings to reflect actual patient experiences rather than controlled experimental conditions. In addition, the mixed-method approach, combining quantitative assessment with qualitative exploration, provides a more comprehensive understanding of patient perspectives and enables deeper insight into the underlying reasons for observed outcomes.
The favorable patient experiences observed in this study are consistent with previous research on wearable monitoring in hospital settings. Weenk et al. reported that wearable monitoring was generally well received by patients and healthcare professionals [10]. Previous studies have also described the importance of comfort, limited interference with daily activities, and perceived benefits for patient acceptability [7,17]. The present study examined these experiences across several domains, including comfort, sleep, reassurance, and willingness to reuse the technology, during routine clinical implementation.
Acceptance scores differed between the Healthdot and viQtor monitoring contexts across several domains, including comfort, sleep quality, reassurance, and willingness to reuse. These findings are consistent with a possible role for device characteristics, but alternative explanations related to ward context, patient population, workflow, and communication cannot be excluded. Factors such as device size, attachment method, and positioning may influence comfort and usability during daily activities and sleep. These findings are consistent with previous research indicating that ergonomic design and wearability are key determinants of user acceptance in wearable technologies [7,17].
Despite the overall high acceptance, qualitative findings revealed variability in patient experiences, which was consistent with the quantitative results. Some participants described tightness, pressure, or skin irritation, particularly in relation to the upper-arm monitoring context; this observation should be interpreted descriptively because the qualitative material was limited and ward and device were linked. The exploratory context differences cannot be separated from ward-level and clinical-context differences. In addition, both qualitative and quantitative findings indicated that patient information and communication remain important areas for improvement. Several patients indicated limited understanding of the device’s purpose and lack of feedback on measured values. This is clinically relevant, as patient understanding may influence perceived benefit, engagement, and acceptance. Previous studies in digital health similarly emphasize the importance of patient education, communication, and organizational factors for successful implementation [18–20].
Reassurance scores were more moderate compared to other domains, suggesting that reassurance from continuous monitoring may not always be consciously perceived. Qualitative findings support this interpretation, as some patients described the sense of being monitored as reassuring, whereas others did not report an additional sense of reassurance.
No statistically significant associations were observed between overall acceptance and the measured patient characteristics. However, the study was not designed or powered to determine whether monitoring context was more important than demographic or clinical characteristics.
Several limitations should be considered. First, device and ward were almost completely confounded, preventing separation of device effects from ward-level and clinical-context effects. Second, the questionnaire was adapted from TAM2 but was not formally piloted or validated in this population. Third, the qualitative component was limited to two semistructured interviews and opportunistic free-text responses; no formal saturation assessment, codebook, or inter-coder reliability assessment was used. Fourth, the cross-sectional design captures experiences at a single time point and does not reflect changes over time. Fifth, the single-center design and voluntary participation may limit generalizability and may have introduced selection and social-desirability bias.The generally favorable patient-reported experiences support continued evaluation of wearable monitoring in surgical care. Before broader implementation claims can be made, questionnaire validity, clinical effectiveness, safety, cost-effectiveness, and longer-term implementation outcomes should be evaluated. The reported variability in comfort and patient information also supports the use of clear communication and structured patient education during implementation.
Patients in this single-center cohort generally reported favorable experiences with continuous wearable vital-sign monitoring. Acceptance differed between the two monitoring contexts, but because device and ward were almost completely linked, these differences cannot be attributed to device type. The findings highlight the importance of comfort, patient information, and communication. Further studies should validate the questionnaire and evaluate clinical effectiveness, safety, cost-effectiveness, and longer-term implementation outcomes.
Supporting information
S2 Table. Full regression and sensitivity analyses.
https://doi.org/10.1371/journal.pone.0363253.s002
(DOCX)
S1 Dataset. De-identified participant-level quantitative dataset.
https://doi.org/10.1371/journal.pone.0363253.s003
(XLSX)
S2 Dataset. English-language, de-identified qualitative dataset comprising translated questionnaire free-text responses, respondent-level category codes, category definitions, and selected paraphrased excerpts from contemporaneous interview notes.
https://doi.org/10.1371/journal.pone.0363253.s004
(XLSX)
Acknowledgments
The authors thank the nursing staff of the participating surgical wards at Catharina Hospital Eindhoven for their support. We also acknowledge SmartQare (Eindhoven, the Netherlands) for providing the wearable monitoring devices.
References
- 1. Brown H, Terrence J, Vasquez P, Bates DW, Zimlichman E. Continuous monitoring in an inpatient medical-surgical unit: A controlled clinical trial. Am J Med. 2014;127(3):226–32. pmid:24342543
- 2. Downey CL, Tahir W, Randell R, Brown JM, Jayne DG. Strengths and limitations of early warning scores: A systematic review and narrative synthesis. Int J Nurs Stud. 2017;76:106–19. pmid:28950188
- 3. Escobar GJ, Liu VX, Schuler A, Lawson B, Greene JD, Kipnis P. Automated identification of adults at risk for in-hospital clinical deterioration. N Engl J Med. 2020;383(20):1951–60. pmid:33176085
- 4. Hernandez-Silveira M, Ahmed K, Ang SS. Wearable sensors for monitoring the physiological and biochemical profile of the patient. J Clin Monit Comput. 2015;29:367–75.
- 5. Joshi M, Ashrafian H, Aufegger L, Khan S, Arora S, Cooke G, et al. Wearable sensors to improve detection of patient deterioration. Expert Rev Med Devices. 2019;16(2):145–54. pmid:30580650
- 6. Michard F, Kalkman CJ, Smulders YM. Performance of wearable sensors for continuous vital sign monitoring in hospitalized patients. J Clin Monit Comput. 2021;35:1235–43.
- 7. Weenk M, van Goor H, Frietman B, Engelen LJ, van Laarhoven CJ, Smit J, et al. Continuous monitoring of vital signs using wearable devices on the general ward: Pilot study. JMIR Mhealth Uhealth. 2017;5(7):e91. pmid:28679490
- 8. Subbe CP, Kruger M, Rutherford P, Gemmel L. Validation of a modified Early Warning Score in medical admissions. QJM. 2001;94(10):521–6. pmid:11588210
- 9. Jerry EE, Bouwman ARA, van Esch EMG, Richardson R, Nienhuijs SW. Remote monitoring with a reusable device upon implementation on a surgical department (REQUEST trial). World J Surg. 2026;50(3):676–83. pmid:41672938
- 10. Weenk M, Bredie SJ, Koeneman M, Hesselink G, van Goor H, van de Belt TH. Continuous monitoring of vital signs in the general ward using wearable devices: Randomized controlled trial. J Med Internet Res. 2020;22(6):e15471. pmid:32519972
- 11. Breteler MJM, KleinJan EJ, Dohmen DAJ, Leenen LPH, van Hillegersberg R, Ruurda JP, et al. Vital signs monitoring with wearable sensors in high-risk surgical patients: A clinical validation study. Anesthesiology. 2020;132(3):424–39. pmid:31743149
- 12. Leenen JPL, Dijkman EM, van Dijk JD, van Westreenen HL, Kalkman C, Schoonhoven L, et al. Feasibility of continuous monitoring of vital signs in surgical patients on a general ward: An observational cohort study. BMJ Open. 2021;11(2):e042735. pmid:33597138
- 13. Cuschieri S. The STROBE guidelines. Saudi J Anaesth. 2019;13(Suppl 1):S31–4. pmid:30930717
- 14. O’Brien BC, Harris IB, Beckman TJ, Reed DA, Cook DA. Standards for reporting qualitative research: A synthesis of recommendations. Acad Med. 2014;89(9):1245–51. pmid:24979285
- 15. Venkatesh V, Davis FD. A theoretical extension of the Technology Acceptance Model: Four longitudinal field studies. Manag Sci. 2000;46(2):186–204.
- 16. Braun V, Clarke V. Using thematic analysis in psychology. Qual Res Psychol. 2006;3(2):77–101.
- 17. Sprogis SK, Currey J, Considine J. Patient acceptability of wearable vital sign monitoring technologies in the acute care setting: A systematic review. J Clin Nurs. 2019;28(15–16):2732–44. pmid:31017338
- 18. Greenhalgh T, Wherton J, Papoutsi C, Lynch J, Hughes G, A’Court C, et al. Beyond adoption: A new framework for theorizing and evaluating nonadoption, abandonment, and challenges to the scale-up, spread, and sustainability of health and care technologies. J Med Internet Res. 2017;19(11):e367. pmid:29092808
- 19. Khalil V. Barriers and enablers of implementing new technology in healthcare: A perspective for healthcare providers. Curr Opin Epidemiol Public Health. 2025;4(4):56–61.
- 20. Suggs LS. A 10-year retrospective of research in new technologies for health communication. J Health Commun. 2006;11(1):61–74. pmid:16546919